You are here: Home Lab Products Autoclave Autoclave 16NE

Autoclave 16NE
View Full-Size Image


Autoclave 16NE

Product Description:

autoclave16ne1

Model

power supply

V/Hz

Power

W

Chamber Sizemm

Packing size(mm)

(Net.kg)

(Gross.kg)

STE-12L-E

110V-240V/50Hz-60Hz

1600W

Φ200*360

690*560*490

38

43

STE-16L-E

110V-240V/50Hz-60Hz

1600W

Φ230*360

690*560*490

39

44

1 instrument rack and 3 trays

- Sterilization temperature: 121 and 134 degrees.

- 3 buttons, very easy to use, any people can run it easily.

- All cycles are fully automatic; there is no need for human intervention in cycle management.

Thermal fractionated gravity system: STE-16E-N works with air dilution by using a series of steam pulses, in which the chamber is alternately pressurized and then depressurized to near atmospheric pressure. All these are achieved thanks to a high performance steam generator.

Program Temperature (˚c) Pressure (bar) Sterilization Time(Min) Total Time (Min)

Easy-to-read: bright blue LCD panel highlights pressure, temperature, time, and process alerts, working conditions of other functions, such as if connection status of printer, water level in tanks.

Fill and drain tanks: TINGET autoclaves are fitted with integral overhead type clean water tank that can be manually filled and cleaned very easily, they are also fitted with an inner waste tanks. Each tank has sensors to alert you when empty or full. Both tanks have quick connections for and draining.

Malfunction detection system: An audible alarm will alert you to any cycle abnormality, and also an error code will be shown on LCD display.

Separate steam generator: A major feature of this product is its innovative steam generator. This design has many advantages. Steam is generated quickly, helping to reduce overall cycle times. There are no heaters in the bottom of the chamber ensuring that the chamber is clean and unencumbered at all times. This in turn helps to reduce the risk of cross-contamination.

Pressure protection locking device (optional): This is not an standard feature for this promtion model STE-16EN, but it is optional.

Thermal drying phase: For Class N autoclave, a nearly completely drying can be achieved by thermal outward; a drying phase is preset into all pre-programmed cycles that automatically runs immediately after sterilization phase.

Documentation:

1. Printer: An external data printer is optional all models of second generation autoclaves

2. USB port: (optional)!!!

3. Automatic storing data: Sterilization data of last 20 cycles can be automatically stored in TINGET autoclave system, they can be chosen to print out or store in USB stick at any time.

Pre-programmed cycles: 7 pre-programmed cycles are installed onto each model. See following table for details. All cycle data, status and messages are clearly displayed on an LCD screen. The sterilization time and dry time can be adjusted by user own requirements in order to obtain best sterilization effects.

PROGRAM

Temperature(˚c)

Pressure(bar)

Sterilization Time(min)

Total Time(min)

SOLID 123

134

2.1

4

18-25

SOLID 121

121

1.1

20

30-38

WRAPPED 134

134

2.1

6

30-35

WRAPPED 121

121

1.1

20

35-45

TEXTILE 134

134

2.1

10

45-55

TEXTILE 121

121

1.1

30

50-60

PRION

134

2.1

18

50-60

LIQUID 134

134

2.1

3.5

45-55

LIQUID 121

121

1.1

3.5

55-70


Autoclave 16NE